Public Health Unit Infection Prevention and Control Lapse Report
Initial Report
Premises/Facility Under Investigation
Casselman Medical Clinic
629 Principale Street
Casselman, Ontario
K0A 1M0
Type of Premises/Facility
Medical Clinic
Date Board of Health Became Aware of IPAC Lapse
October 13th, 2026
Date IPAC lapse was linked to the premise/facility
October 13th, 2026
Date of Initial Report Posting
November 16th, 2026
How the IPAC Lapse was Identified
Clinic requested IPAC education and support
Summary Description of the IPAC Lapse
During the inspection, deficiencies in the reprocessing of medical devices were identified.
The autoclave and reprocessing procedures were not documented and staff were unable to demonstrate the required reprocessing practices.
The items were cleaned with an enzymatic cleaning product that was not used in accordance with the manufacturer's instructions.
No biological indicators were noted to be in use.
IPAC Lapse Investigation
Issue referred to regulatory college and/or other stakeholder notified?
Yes
If the Lapse Involved a Member of a Regulatory College, Was the Issue Referred to that Regulatory College?
College of Physicians and Surgeons of Ontario
Were any Corrective Measures Recommended and/or Implemented?
Yes
Further Details/Steps
•Cease all reprocessing activities until the sterilizer has undergone required testing and maintenance.
•Remove all previously sterilized equipment from circulation and use single-use items only.
•Complete spore testing of the autoclave and arrange for the autoclave to be serviced, including repair of the printer.
If you have any further questions, please contact:
Geneviève Lord
Program Manager / Gestionnaire de programmes
glord@eohu.ca
613-933-1375 x 1233